Technical data
Performance Tests and Calibration 4
34405A User’s and Service Guide 77
AC Voltage Verification Test
Configuration: AC Volts (CONFigure[:VOLTage]:AC)
1 Select the AC Voltage function.
2 Select each range in the order shown below. Provide the indicated input
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
[1] Select 5½- digit measurement resolution
Q = Quick performance verification test points
Table 9 AC Volts Verification Test
Vrms Input Frequency Function
[1]
Range Quick Check
Error from Nominal
1 year
100mV 1kHz AC Voltage 100mV ± 0.3 mV
100mV 30kHz 100mV ± 1.8 mV
100mV 100kHz 100mV ± 5.3 mV
1V 1kHz 1V Q ± 3.0 mV
1V 30kHz 1V ± 11 mV
1V 100kHz 1V ± 32 mV
10V 45Hz 10V ± 110 mV
10V 1kHz 10V ± 30 mV
10V 30kHz 10V Q ± 0.11 V
10V 100kHz 10V ± 0.32 V
100V 1kHz 100V Q ± 0.3 V
100V 30kHz 100V ± 1.1 V
100V 100kHz 100V ± 3.2 V
750V 1kHz 750V ± 2.25 V
Caution: Set the calibrator output to 0V before disconnecting it from the multimeter input terminals.