Technical data

66 34405A User’s and Service Guide
4 Performance Tests and Calibration
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below.If the exact instrument is not
available, substitute calibration standards of equivalent accuracy.
A suggested alternate method would be to use the Agilent 3458A 8½
- Digit Digital Multimeter to measure less accurate yet stable sources. The
output value measured from the source can be entered into the instrument
as the target calibration value.
Table 3 Recommended Test Equipment
Application Recommended Equipment Recommended Accuracy Requirements
Zero Calibration Shorting Plug--Dual banana plug with copper wire
short between the two terminals
DC Voltage Fluke 5520A <1/5 instrument 1 year spec
DC Current Fluke 5520A <1/5 instrument 1 year spec
Resistance Fluke 5520A <1/5 instrument 1 year spec
AC Voltage Fluke 5520A <1/5 instrument 1 year spec
AC Current Fluke 5520A <1/5 instrument 1 year spec
Frequency Fluke 5520A <1/5 instrument 1 year spec
Capacitance Fluke 5520A <1/5 instrument 1 year spec