Technical data

Features and Functions 2
34405A User’s and Service Guide 41
Storing and Recalling Instrument States
You can save and recall complete instrument states including all front
panel settings, all math registers, all Utility Menu settings, and all bus
specific settings. There are four user storage registers numbered 1 through
4. An additional state, state 0, is managed by the instrument and stores
the last power- down state. The instrument automatically saves the
complete instrument configuration to State 0 whenever a power- down
event occurs.
For remote operation, refer to the MEMory Subsystem, the *SAV, and *RCL
commands in the Agilent 34405A Online Programmer’s Reference online
help.
Storing a State
Before storing an instrument state, select the measurement function,
range, math operations, and so on, that you want saved as a state. To
store the instrument state:
1 Press , the display Store and Recall annunciators will begin
flashing.
2 Press or until only the Store annunciator is flashing.
3 Press again.
4 Press or until the state number (1- 4) you want to use is
shown flashing in the secondary display.
5 Press to store the state. The secondary display briefly shows
donE when the state is successfully saved.
NOTE
The store function in the utility menu must be enabled (On) before you can store states.
Refer to “Using the Utility Menu” on page 36 for details.
Store
Recall
Store
Recall
Store
Recall
Store
Recall
Store
Recall
Store
Recall