General Characteristics
Table 30 General Characteristics
Power Supply
• 100V/120V(127V)/220V(230V)/240V ± 10%
• AC line frequency 45Hz - 66Hz and (360Hz - 440Hz, 100/120V operation)
Power Consumption
• 16VA maximum, <11W average
Operating Environment
• Full accuracy at 0° C to 55° C
• Full accuracy to 80% RH at 30° C (non-condensing)
• Altitude up to 3000 meters
Storage Compliance
• - 40° C to 70° C
Safety Compliance
• Certified by CSA for IEC/EN/CSA/UL 61010-1 2
nd
Edition
Measurement Category
• CAT II, 300V: CAT I 1000Vdc, 750Vac rms, 2500Vpk transient over voltages
• Pollution degree 2
EMC Compliance
• Certified to IEC61326-1: 2005 / EN61326-1:2006
• CISPR 11:2003 / EN 55011:2007 Group 1 Class A
• Canada: ICES-001:2004
• Australia/New Zealand: AS/NZS CISPR 11:2004
Shock and Vibration
• Tested to IEC/EN 60086-2
Dimension (HxWxD)
• Rack: 88.5mm x 212.6mm x 272.3mm