Technical data

132 34405A User’s and Service Guide
6 Specifications
Crest Factor:
Maximum 5:1 at full scale
Input Impedance:
1MΩ ± 2% in parallel with < 100pF of all ranges
Input Protection:
750V rms on all ranges (HI terminal)
AC Current
Measurement Method:
DC coupled to the fuse and current shunt, AC coupled true rms measurement (measures the
AC component only)
Shunt Resistance:
0.1Ω to 10Ω for 10mA to 1.2A range
0.01Ω for 12A range
Input Protection:
Externally accessible 1.25A, 500V FH fuse for I terminal
Internal 15A, 600V fuse for 12A terminal
Frequency
Measurement Method:
Reciprocal counting technique. AC coupled input using AC voltage function.
Signal Level:
10% of range to full scale input on all ranges
Auto or manual range selection
Gate Time:
0.1 second or 1 period of the input signal, whichever is longer.
Input Protection:
750V rms on all ranges (HI terminal)
Table 29 Supplemental Measurement Specifications