Technical data

Specifications 6
34405A User’s and Service Guide 131
Continuity / Diode Test
Measurement Method:
Uses 0.83mA ± 0.2% constant current source, < 5V open circuit voltage
Response Time:
70 samples/second with audible tone
Continuity Threshold:
10Ω fixed
Input Protection:
1000V (HI terminal)
Temperature
Measurement Method:
2-wire Ohms measurement of 5kΩ thermistor sensor (YSI 4407) with computer conversion
Auto-ranging measurement, no manual range selection
Input Protection:
1000V (HI terminal)
Measurement Noise Rejection
CMR (Common Mode Rejection) For 1kΩ unbalance LO lead
DC 120 dB
AC 70 dB
NMR (Normal Mode Rejection) For 60 Hz (50 Hz) ± 0.1%
5½ digits 65 dB (55 dB)
4½ digits 0 dB
AC Voltage
Measurement Method:
AC coupled true-rms - measure the AC component with up to 400 VDC bias any range
Table 29 Supplemental Measurement Specifications