Technical data
126 34405A User’s and Service Guide
6 Specifications
AC Specifications
[1]
WARNING
Exceeding the crest factor limit may result in an incorrect or a lower
reading. Do not exceed the crest factor limit to avoid instrument
damage and the risk of electric shock.
Table 25 AC Accuracy ± (% of reading + % of range)
Function Range
[5]
Frequency
1 Year
23° C ± 5° C
Temperature Coefficient
0° C - 18° C
28° C - 55° C
True RMS AC Voltage
[6]
100.000 mV 20 Hz - 45 Hz 1+0.1 0.02+0.02
45 Hz - 10 kHz 0.2+0.1 0.02+0.02
10 kHz - 30 kHz 1.5+0.3 0.05+0.02
30 kHz - 100 kHz
[7]
5.0+0.3 0.10+0.02
1.00000 V to 750.00 V 20 Hz - 45 Hz 1+0.1
(14)
0.02+0.02
45 Hz - 10 kHz 0.2+0.1 0.02+0.02
10 kHz - 30 kHz 1+0.1 0.05+0.02
30 kHz - 100 kHz
(7)
3+0.2
(15)
0.10+0.02
True RMS AC Current
[8]
10.0000 mA to 10.0000 A 20 Hz - 45 Hz 1.5+0.1 0.02+0.02
45 Hz - 1 kHz 0.5+0.1 0.02+0.02
1 kHz - 10 kHz
[9]
2+0.2 0.02+0.02