Technical data
Performance Tests and Calibration 4
34405A User’s and Service Guide 101
Finishing the Adjustments
1 Remove all shorting plugs and connections from the instrument.
2 Reset the Calibration Message (see below).
3 Record the new Calibration Count (see page 102).
4 Press and simultaneously to exit the Adjustment Mode.
The instrument will be secured and returns to DC Voltage, autoranging
measurements.
Table 21 Capacitance Gain Adjustment
Input Function Cal Item
Input terminals open (remove any test leads or
shorting plugs from the input terminals)
Capacitance Open
0.4 nF 0.4 nF
1 nF 1 nF
10 nF 10 nF
100 nF 100 nF
1 μF1 μF
10 μF10 μF
100 μF100 μF
1000 μF 1000 μF
10000 μF 10000 μF
Shift
DCV