Technical data

Performance Verification Tests
You can perform three different levels of performance verification tests:
Self-Test A series of internal verification tests that give a high
confidence that the multimeter is operational.
Quick Verification A combination of the internal self-tests and
selected verification tests.
Performance Verification Tests An extensive set of tests that are
recommended as an acceptance test when you first receive the
multimeter or after performing adjustments.
Self-Test
A brief power-on self-test occurs automatically whenever you turn on
the multimeter. This limited test assures that the multimeter is capable
of operation.
To perform a complete self-test, including over 25 tests, hold down
the
Shift key as you press the Power switch to turn on the multimeter;
hold down the key for more than 5 seconds (a complete description of
these tests can be found in chapter 6.) The multimeter meter will
automatically perform the complete self-test procedure when you
release the key. The self-test will complete in approximately 20 seconds.
You can perform many tests individually (or all tests at once) using the
TEST command in the SYS MENU. You can also perform a self-test from
the remote interface. For further information, see chapter 3 in the
Agilent 34401A User’s Guide.
If the self-test is successful,PASS is displayed on the front panel.
If the self-test fails, “FAIL” is displayed and the ERROR annunciator
turns on. If repair is required, see chapter 6, “Service,” for further details.
If all tests pass, you can have a high confidence (90%) that the
multimeter is operational.
4
Chapter 4 Calibration Procedures
Performance Verification Tests
65