Technical data
Performance Verification Tests
You can perform three different levels of performance verification tests:
• Self-Test A series of internal verification tests that give a high
confidence that the multimeter is operational.
• Quick Verification A combination of the internal self-tests and
selected verification tests.
• Performance Verification Tests An extensive set of tests that are
recommended as an acceptance test when you first receive the
multimeter or after performing adjustments.
Self-Test
A brief power-on self-test occurs automatically whenever you turn on
the multimeter. This limited test assures that the multimeter is capable
of operation.
To perform a complete self-test, including over 25 tests, hold down
the
Shift key as you press the Power switch to turn on the multimeter;
hold down the key for more than 5 seconds (a complete description of
these tests can be found in chapter 6.) The multimeter meter will
automatically perform the complete self-test procedure when you
release the key. The self-test will complete in approximately 20 seconds.
You can perform many tests individually (or all tests at once) using the
TEST command in the SYS MENU. You can also perform a self-test from
the remote interface. For further information, see chapter 3 in the
Agilent 34401A User’s Guide.
• If the self-test is successful, “PASS” is displayed on the front panel.
• If the self-test fails, “FAIL” is displayed and the ERROR annunciator
turns on. If repair is required, see chapter 6, “Service,” for further details.
• If all tests pass, you can have a high confidence (90%) that the
multimeter is operational.
4
Chapter 4 Calibration Procedures
Performance Verification Tests
65