Technical data
Calibration Procedures
• Agilent Calibration Services 61
• Calibration Interval 61
• Time Required for Calibration 61
• Automating Calibration Procedures 62
• Recommended Test Equipment 63
• Test Considerations 64
• Performance Verification Tests 65
• Zero Offset Verification 67
• Gain Verification 69
• Optional AC Performance Verification Tests 72
• Calibration Security Code 73
• Calibration Count 75
• Calibration Message 75
• Calibration Procedures 76
• Aborting a Calibration in Progress 76
• Zero Adjustment 77
• Gain Adjustment 79
• Optional Gain Calibration Procedures 82
• Understanding the AC Signal Filter 85
• Understanding Resolution 86
• Error Messages 89
The performance verification tests use the multimeter’s specifications
listed in chapter 1, “Specifications,” starting on page 11.
Closed-Case Electronic Calibration The multimeter features
closed-case electronic calibration since there are no internal mechanical
adjustments required. The multimeter automatically prompts you for
the required full-scale input when performing a range calibration.
The multimeter measures the applied input and calculates correction
factors based upon the input reference value you specify. The new
correction factors are stored in non-volatile memory until the next
calibration adjustment is performed. (Non-volatile memory does not
change when power has been off or after a remote interface reset.)
60