Technical data

Calibration Procedures
Agilent Calibration Services 61
Calibration Interval 61
Time Required for Calibration 61
Automating Calibration Procedures 62
Recommended Test Equipment 63
Test Considerations 64
Performance Verification Tests 65
Zero Offset Verification 67
Gain Verification 69
Optional AC Performance Verification Tests 72
Calibration Security Code 73
Calibration Count 75
Calibration Message 75
Calibration Procedures 76
Aborting a Calibration in Progress 76
Zero Adjustment 77
Gain Adjustment 79
Optional Gain Calibration Procedures 82
Understanding the AC Signal Filter 85
Understanding Resolution 86
Error Messages 89
The performance verification tests use the multimeter’s specifications
listed in chapter 1, “Specifications,” starting on page 11.
Closed-Case Electronic Calibration The multimeter features
closed-case electronic calibration since there are no internal mechanical
adjustments required. The multimeter automatically prompts you for
the required full-scale input when performing a range calibration.
The multimeter measures the applied input and calculates correction
factors based upon the input reference value you specify. The new
correction factors are stored in non-volatile memory until the next
calibration adjustment is performed. (Non-volatile memory does not
change when power has been off or after a remote interface reset.)
60