Specifications

Power Meter Remote Operation
Using Sensor Calibration Tables
Agilent E4418B/E4419B Programming Guide 1-35
Each string parameter returned indicates the name, type and size of a
stored sensor calibration table:
<string>,<type>,<size>
The <string>, <type> and <size> are all character data. The
<type> is always TABL. The <size> is displayed in bytes.
For example, a sample of the response may look like:
560,8020,“Sensor_1,TABL,220”,”Sensor_2,TABL,340” ....
The power meter is shipped with a set of predefined sensor calibration
tables. The data in these sensor calibration tables is based on statistical
averages for a range of Agilent Technologies Power Sensors (see Chapter
2, Editing Sensor Calibration Tables in the Users Guide). These power
sensors are:
DEFAULT
1
Agilent 8481A
Agilent 8482A
2
Agilent 8483A
Agilent 8481D
Agilent 8485A
Agilent R8486A
Agilent Q8486A
Agilent R8486D
Agilent 8487A
For further information on naming sensor calibration tables see Naming
Sensor Calibration Tables, on page 1-36.
1. DEFAULT is a sensor calibration table in which the reference calibration
factor and calibration factors are 100%. This sensor calibration table can be
used during the performance testing of the power meter.
2. The Agilent 8482B and Agilent 8482H power sensors use the same data as
the Agilent 8482A.