*DDT <arbitrary block program data>|<string program data>
15-8 Agilent E4418B/E4419B Programming Guide
*DDT <arbitrary block program data>|<string program data>
The *DDT (Define Device Trigger) command determines the power meters
response to a GET (Group Execute Trigger) message or *TRG common
command. This command effectively turns GET and *TRG into queries,
with the measured power being returned.
Syntax
Parameters
Examples of <arbitrary block program data> parameters are:
#15FETC? and #206FETCh?
Examples of <string program data> are:
"FETCh1?", "FETCh?" and "TRIG1;FETC1"
Reset Condition
On reset, the <action> field of *DDT is set to *TRG.
Type Description Range of Values
arbitrary block program
data
The command which is
executed on a GET or
*TRG.
#nN<action>
1,2
1. The <action> field of the parameter may contain:
FETC?
FETC1?
FETC2? (Agilent E4419B only)
*TRG
TRIG1
TRIG2 (Agilent E4419B only)
2. The first digit after the # indicates the number of following digits. The following
digits indicate the length of the data.
string program data
<action>
1
string program data
*DDT
Space
arbitrary block program data
?