Specifications
SYSTem Subsystem
SYSTem:PRESet
Agilent E4418B/E4419B Programming Guide 11-29
SYSTem:PRESet
This command is used to preset the power meter. The result of this
command is similar to the *RST command. Refer to Table 11-1 for more
information about the preset configuration.
Note All settings apply to both *RST and to SYSTem:PREset unless
stated otherwise.
Syntax
Example
Table 11-1: Preset Settings
SYST:PRES This command presets the
power meter.
Command Setting Comments
CALC:GAIN[:MAGN] 0.000 dB Display offset value
CALC:GAIN:STAT OFF Display offset disabled
CALC:LIM:CLE:AUTO ON Clear limit data at INIT
CALC:LIM:FAIL OFF No failures
CALC:LIM:FCO OFF Limit failure count is zero
CALC:LIM:LOW[:DATA] -90 dBM or -90 dB Lower limit
CALC:LIM:STAT OFF Window limits checking disabled
CALC:LIM:UPP[:DATA] +90 dBM or +90 dB
CALC:MATH[:EXPR] Agilent E4418B:
Upper - channel A
Lower - channel A
Agilent E4419B
Upper - channel A
Lower - channel B
Math expression
CALC:REL[:MAGN]:AUTO OFF Reference value disabled
CALC:REL:STAT OFF Relative offset disabled
CAL:AUTO OFF Automatic calibration disabled
SYST
:PRES
prog.book Page 29 Thursday, June 7, 2001 2:55 PM