Specifications

STATus Subsystem
STATus:OPERation:CALibrating[:SUMMary]
10-12 Agilent E4418B/E4419B Programming Guide
STATus:OPERation:CALibrating[:SUMMary]
The operation status calibrating summary register set contains
information on the calibrating status of the power meter.
The following bits in these registers are used by the power meter:
These bits are set at the beginning of zeroing (CALibration:ZERO:AUTO
ONCE) and at the beginning of calibration (CALibration:AUTO ONCE).
Also for the compound command/query CALibration[:ALL]?, this bit is
set at the beginning of the calibration sequence.
These bits are cleared at the end of zeroing or calibration.
Syntax
Bit
Number
Decimal
Weight
Definition
0-Not used
1 2 Channel A CALibrating Status
2 4 Channel B CALibrating Status (Agilent E4419B only)
3-15 - Not used
STAT
:OPER
:CAL :SUMM