Specifications

SENSe Subsystem
[SENSe[1]]|SENSe2:LIMit:FCOunt?
9-44 Agilent E4418B/E4419B Programming Guide
[SENSe[1]]|SENSe2:LIMit:FCOunt?
This query is used to return the total number of limit failures for the
specified channel.
If the appropriate STATe commands are set to ON, each time a
measurement is initiated on the specified channel and the measurement
result is outside the limits, the counter is incremented by one.
If the measured value is equal to a limit, this is a limit pass.
The counter is reset to zero by any of the following commands:
*RST
[SENSe[1]]|SENSe2:LIMit:CLEar:IMMediate
[SENSe[1]]|SENSe2:LIMit:CLEar:AUTO ON
When [SENSe[1]]|SENSe2:LIMit:CLEar:AUTO is set to ON, the
counter is set to zero each time a measurement:
is measured using MEASure?
is read using READ?
is initiated using:
INITiate[:IMMediate] or,
INITiate:CONTinuous ON.
When [SENSe[1]]|SENSe2:LIMit:CLEar:AUTO is set to ONCE, the
counter is set to zero the first time a measurement:
is measured using MEASure?
is read using READ?
is initiated using:
INITiate[:IMMediate] or,
INITiate:CONTinuous ON.
The maximum number of errors is 2
16
-1. If more than 2
16
-1 errors are
detected the counter returns to zero.