Specifications

SENSe Subsystem
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed] <numeric_value>
Agilent E4418B/E4419B Programming Guide 9-37
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed] <numeric_value>
This command is used to enter a frequency. If the frequency does not
correspond directly to a frequency in the sensor calibration table, the
power meter calculates the calibration factor using linear interpolation.
For Agilent 8480 series power sensors the power meter uses linear
interpolation to calculate the calibration factor for the frequency entered if
[SENSe[1]]|SENSe2:CORRection:CSET:STATe is ON. For
Agilent E-series power sensors, the calibration factor is calculated using
the sensors calibration factor data held in its EEPROM.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value A numeric value for the frequency. The
default value, DEF, is 50 MHz. The value
for MIN is 1 kHz. The value for MAX is
999.999 GHz. The default units are HZ.
1kHz to
999.999 GHz
1
DEF
MIN
MAX
1. The following measurement units can be used:
Hz
kHz (10
3
)
MHz (10
6
)
GHz (10
9
)
:CW
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX