Specifications

SENSe Subsystem
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2[:SELect] <string>
9-16 Agilent E4418B/E4419B Programming Guide
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect]?
The name of the selected table is returned as a quoted string. If no table is
selected an empty string is returned.
Query Example
Error Messages
If <string> is not valid, error -224, Illegal parameter value occurs.
If a table called <string> does not exist, error -256, File name not
found occurs.
When a sensor calibration table is selected, the power meter verifies that
the number of calibration points defined is one more than the number of
frequency points defined. When a frequency dependent offset table is
selected, the power meter verifies that the number of offset points defined
is equal to the number of frequency points defined. If this is not the case,
error -226, Lists not the same length occurs.
If the CSET1 command is used when an Agilent E-series power sensor is
connected the error -241, Hardware missing occurs.
CORR:CSET1? This command queries the sensor
calibration table currently used for
channel A.