Specifications

MEMory Subsystem
MEMory:CATalog:TABLe?
7-8 Agilent E4418B/E4419B Programming Guide
The power meter is shipped with a set of predefined sensor calibration
tables. The data in these sensor calibration tables is based on statistical
averages for a range of Hewlett-Packard power sensors. These tables can
be altered. The predefined data is listed in your Users Guide. These power
sensors are:
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9 and ten frequency dependent offset tables named
CUSTOM_A through CUSTOM_J which do not contain any data when the
power meter is shipped from the factory.
Syntax
Example
Table Power Sensor Table Name
0NoneDEFAULT
1
1. Default is a sensor calibration table in which the reference calibration
factor and calibration factors are 100%. This sensor calibration table can be
used during the performance testing of the power meter.
1 Agilent 8481A HP8481A
2 Agilent 8482A, 8482B, 8482H HP8482A
3 Agilent 8483A HP8483A
4 Agilent 8481D HP8481D
5 Agilent 8485A HP8485A
6 Agilent R8486A R8486A
7 Agilent Q8486A Q8486A
8 Agilent R8486D R8486D
9 Agilent 8487A HP8487A
MEM:CAT:TABL? This command queries the list of
stored tables.
MEM
:CAT
:TABL ?
prog.book Page 8 Thursday, June 7, 2001 2:55 PM