Specifications
CALibration Subsystem
CALibration[1|2]:RCALibration <Boolean>
4-10 Agilent E4418B/E4419B Programming Guide
CALibration[1|2]:RCALibration <Boolean>
This command enables and disables the zero/cal lockout facility. With the
lockout facility enabled the power meter is prevented from making
measurements until the sensor connected has been zeroed and calibrated.
Syntax
Example
Reset Condition
On reset, the state of the zero/cal lockout is unaffected.
Query
CALibration[1|2]:RCALibration?
The query enters a 1 or 0 into the output buffer indicating whether
zero/cal lockout is enabled or disabled.
• 1 is returned if zero/cal lockout is enabled.
• 0 is returned if zero/cal lockout is disabled.
Query Example
CAL1:RCAL 1 This command enables the zero
/cal lockout facility on channel A.
CAL1:RCAL? This command queries whether or
not the zero/cal lockout facility is
enabled for channel A.
CAL
1
2
:RCAL
0|OFF
1|ON
Space
?