Specifications
MEASurement Instructions
READ[1|2][:SCALar][:POWer:AC]? [<expected_value>[,<resolution>[,<source list>]]]
Agilent E4418B/E4419B Programming Guide 2-37
Example
Error Messages
INITiate:CONTinuous must be set to OFF, otherwise error -213, “INIT
ignored” occurs.
If TRIGger:SOURce is set to BUS or HOLD, error -214, “Trigger deadlock”
occurs.
If the expected value and resolution parameters are not the same as the
current expected value and resolution settings on the specified window,
error -221, “Settings conflict” occurs.
resolution A numeric value for the resolution.
If it is unspecified the current
resolution setting is used. If a
value is entered it should
correspond to the current
resolution setting otherwise an
error occurs.
1 to 4
2
1.0, 0.1, 0.01, 0.001
DEF
1
source list This is the channel on which the
command is implemented.
If unspecified the current window
setup is used. However, on the
Agilent E4419B, if the window
shows a ratio or difference
measurement, the upper window
defaults to channel A and the
lower window to channel B.
(@1)
(@2) (Agilent E4419B
only)
1. The mnemonic
DEF means DEFault. This is not equivalent to the DEFault
parameter used in the command sub-systems. The parameters must be entered in
the specified order. If parameters are omitted, they will default from the right.
The parameter DEFault is used as a place holder. Specifying
DEF leaves the
parameter value unchanged.
2. When the measurement result is linear this parameter represents the number
of significant digits. When the measurement result is logarithmic 1 to 4
represents a resolution of 1, 0.1, 0.01 and 0.001 respectively.
READ2:POW:AC? This command queries the lower
window’s measurement.
Item Description/Default Range of Values