Specifications

Power Meter Remote Operation
Status Reporting
1-84 Agilent E4418B/E4419B Programming Guide
The condition bits are set and cleared under the following conditions:
Bit Number Meaning EVENts Causing Bit Changes
3 POWer
Summary
This is a summary bit for the Questionable POWer
Register.
SET:
Error -230, Data corrupt or stale
Error -231, Data questionable;Input Overload
Error -231, Data questionable;Input Overload ChA
1
Error -231, Data questionable;Input Overload ChB
1
Error -231, Data questionable;PLEASE ZERO
Error -231, Data questionable;PLEASE ZERO ChA
1
Error -231, Data questionable;PLEASE ZERO ChB
1
Error -231, Data questionable;Lower window log error
1
Error -231, Data questionable;Upper window log error
1
CLEARED: When no errors are detected by the power
meter during a measurement covering the causes given for
it to set.
1. Agilent E4419B only
8 CALibration
Summary
This is a summary bit for the Questionable CALibration
Register.
SET:
These may be caused by
CALibration[1|2]:ZERO:AUTO ONCE or
CALibration[1|2]:AUTO ONCE or
CALibration[1|2][:ALL] or
CALibration[1|2][:ALL]?.
Error -231, Data questionable; ZERO ERROR
Error -231, Data questionable; ZERO ERROR ChA
1
Error -231, Data questionable; ZERO ERROR ChB
1
Error -231, Data questionable; CAL ERROR
Error -231, Data questionable; CAL ERROR ChA
1
Error -231, Data questionable; CAL ERROR ChB
1
CLEARED: When any of the commands listed above
succeed and no errors are placed on the error queue.
9 Power On
Self Test
SET: This bit is set when the power on self test fails.
CLEARED: When the power on self test passes.
prog.book Page 84 Thursday, June 7, 2001 2:55 PM