34134A.book Page 2 Wednesday, March 25, 2009 4:09 PM Contents Inspect the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Control and Connector Identification Information . . . . . Specifications and characteristics . . . . . . . . . . . . . . . . . . . . . Electrical Specifications . . . . . . . . . . . . . . . .
4134A.book Page 3 Wednesday, March 25, 2009 4:09 PM 34134A AC/DC DMM Current Probe The 34134A Current Probe measures low DC or AC from 10 mA to 50 A DC, 40 A AC. This battery-powered clamp-on probe may be used with digital multimeters, voltmeters, or other voltage-measuring instruments. The probe jaws clamp around the conductor under test, allowing current measurements without breaking the circuit.
34134A.book Page 4 Wednesday, March 25, 2009 4:09 PM 34134A AC/DC DMM Current Probe Specifications and characteristics Specifications and characteristics Operating and environmental characteristics of the 34134A Current Probes are shown below. Electrical Specifications Current Range Output Signal (AC/DC) Accuracy* 1 V/A Range (1 mV/mA) DC: 10 mA to 2 A AC: 10 mA to 1.
34134A.
34134A.book Page 6 Wednesday, March 25, 2009 4:09 PM 34134A AC/DC DMM Current Probe Specifications and characteristics Safety Specifications Working voltage Case protection Drop test Mechanical shock Vibration ! 600 V CAT III IP20 per IEC 529 1m according to IEC 68-2-32 100 G: test per IEC 68-2-27 Test per IEC 68-2-6 frequency range 10 Hz to 55 Hz, amplitude: 0.15mm This symbol signifies that the 34134A AC/DC DMM Current Probe is protected by double or reinforced insulation.
34134A.book Page 7 Wednesday, March 25, 2009 4:09 PM 34134A AC/DC DMM Current Probe Operation Operation 1 Connect the probe to the multimeter or other instrument. CAUTION Be sure to observe the polarities: red = positive terminal, black = common terminal. 2 Select the multimeter range which best corresponds to the measured current. For better reading stability, you may use the DMM 2 Volt range and let the probe “warm up” for one minute before zeroing.
34134A.book Page 8 Wednesday, March 25, 2009 4:09 PM 34134A AC/DC DMM Current Probe Calibration Calibration To guarantee that your instrument complies with the factory specifications, we recommend that the 34134A Current Probe be submitted to your local Agilent Service Center at one year intervals for recalibration, or as required by other standards. Repair This probe contains no user serviceable parts. Please return to your local Agilent Service Center for repair.
34134A.book Page 0 Wednesday, March 25, 2009 4:09 PM © Agilent Technologies 1999, 2009 All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws. Restricted Rights Legend. Use, duplication or disclosure by the U.S. Government is subject to restrictions as set forth in subparagraph (c) (1) (ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.
34134A.book Page 1 Wednesday, March 25, 2009 4:09 PM Safety This apparatus has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Measuring Apparatus, and has been supplied in a safe condition. To ensure safe operation and to keep the product safe, the information, cautions, and warnings in this operating manual must be heeded. In addition, note the external markings on the instrument that are described under "Safety Symbols.