Specifications

8 Release Notes
New Control XTP Firmware
The Medalist i3070 08.30p software release has enabled new Control XTP
firmware version 11110111_1 as a backup file. This firmware version is
needed to solve the “T313 Control card timeout waiting for Psync signal
error” encountered if the NEC 2Mx8bit SRAM Control XTP card is used.
Note that no change is needed if the existing xtp.xsvf (110601116_1) is
working fine with the Control XTP cards in the system.
To use Control XTP firmware version 11110111_1:
1 Locate the file C:\Agilent_ICT\lib\xtp.xsvf and rename it
110601116_1.xsvf.
2 Locate the file C:\Agilent_ICT\lib\11110111_1.xsvf and rename it
xtp.xsvf.
3 Boot up the testhead.
4 Launch the Diagnostics application.
5 Select Configuration > Programmable Card Config and check that the Expected
value is 11110111_1.
6 Click Update All (F1) button to program all the Control XTP cards on the
testhead.
7 Reboot the testhead.
Figure 1
Actual version in
ControlXTP cards
Expected version
in xtp.xsvf
Click to update firmware
in all ControlXTP cards