4 Making Measurements
96 Agilent 3000 Series Oscilloscopes User’s Guide
Delay Between Rising Edges
Measures the delay between two waveforms using the rising edges.
Delay Between Falling Edges
Measures the delay between two waveforms using the falling edges. See
Figure 44 on page 96.
Counter (Frequency)
The 3000 Series oscilloscopes have an integrated 5- digit hardware
frequency counter.
The counter operates on the currently selected trigger source and can
measure frequencies from 5 Hz to the bandwidth of the oscilloscope.
Figure 44 Delay Measurements
Delay from channel
1 to channel 2 falling edge
Time
origin
line
50%
Channel 2
Channel 1
50%