Technical data
Table Of Contents
- Agilent 3000 Series Oscilloscopes-At a Glance
- In This Book
- Contents
- Figures
- Tables
- Getting Started
- Step 1. Inspect the package contents
- Step 2. Turn on the oscilloscope
- Step 3. Load the default oscilloscope setup
- Step 4. Input a waveform
- Step 5. Become familiar with the Front Panel Controls
- Step 6. Become familiar with the oscilloscope display
- Step 7. Use Auto-Scale
- Step 8. Compensate probes
- Step 9. Use the Run Control buttons
- Displaying Data
- Capturing Data
- Making Measurements
- Saving, Recalling, and Printing Data
- Oscilloscope Utility Settings
- Specifications and Characteristics
- Service
- Safety Notices
- Index

Capturing Data 3
Agilent 3000 Series Oscilloscopes User’s Guide 85
To play-back/stop the recording
1 In the Sequence menu (Acquire→Sequence→Mode=Play back), select
Operate to play- back or stop the recording.
• — Appears on the menu when not playing-back; press Operate to
start playing-back the recording.
• — Appears on the menu when playing- back; press Operate to stop
recording.
To select continuous or one-time play-back
1 In the Sequence menu (Acquire→Sequence→Mode=Play back), select
Play Mode to toggle between:
• — Continuous play- back.
• — One- time play- back.
To select the interval between played-back frames
1 In the Sequence menu (Acquire→Sequence→Mode=Play back), select
Interval.
2 Turn the Entry
knob to select an interval from 1 ms to 20 s.
To select the start frame
1 In the Sequence menu (Acquire→Sequence→Mode=Play back), select
Start Frame.
2 Turn the Entry
knob to select a number from l to 1000.
To select the current frame
1 In the Sequence menu (Acquire→Sequence→Mode=Play back), select
Current Frame.
2 Turn the Entry
knob to select a number from l to 1000.
To select the end frame
1 In the Sequence menu (Acquire→Sequence→Mode=Play back), select
End Frame.
2 Turn the Entry
knob to select a number from l to 1000.