Service 8
Agilent 3000 Series Oscilloscopes User’s Guide 145
Channel 2
2mV/div ±6mV +0.96 to +1.04
5mV/div ±15mV +0.96 to +1.04
10 mV/div ±30 mV +0.97 to +1.03
20 mV/div ±60 mV +0.97 to +1.03
50 mV/div ±150 mV +0.97 to +1.03
100 mV/div ±300 mV +0.97 to +1.03
200 mV/div ±600 mV +0.97 to +1.03
500 mV/div ±1.5 V +0.97 to +1.03
1 V/div ±2.4 V +0.97 to +1.03
2 V/div ±6.0 V +0.97 to +1.03
5 V/div ±15.0 V +0.97 to +1.03
Table 21 DC Gain Test (continued)
Vertical
Sensitivity
Power
Supply
Setting
VDMM+ VDMM- VScope+ VScope- Calculated
DC Gain
Offset Gain
Test Limits
Table 22 Analog Bandwidth - Maximum Frequency Check
Vin @ 1 MHz Vout @
1MHz
Calculated Gain
@ 1 MHz (Test
Limit = greater
than -3 dB)
Vin @ Max
Freq
Vout @ Max
Freq
Calculated Gain
@ Max Freq
(Test Limit =
greater than -3
dB)
Channel 1
Channel 2
Max frequency: DSO3062A = 60 MHz, DSO3102A = 100 MHz, DSO3152A = 150 MHz, DSO31202A = 200 MHz