Service 8
Agilent 3000 Series Oscilloscopes User’s Guide 131
Procedure
1 Disconnect all cables from the oscilloscope channel inputs.
2 Press Save/Recall
.
3 Select the Storage item in the Save/Recall menu until Setups appears.
4 Select the Default Setup item in the Save/Recall menu.
5 Press the Acquire
front panel button.
6 Select the Mode item in the Acquire menu until “Average” appears.
Cable Assembly (2
required)
50Ω characteristic impedance, BNC (m)
connectors
Agilent 8120-1840
Adapter BNC Tee (m)(f)(f) Agilent 1250-0781
Adapter (2 required) BNC (f) to dual banana Agilent 1251-2277
Figure 52 Default Setup Menu Item
Table 17 Equipment Required for DC Gain Accuracy Test (continued)
Description Critical Specifications Recommended
Model/Part Numbers