8Service
130 Agilent 3000 Series Oscilloscopes User’s Guide
Vertical Performance Verification
This section contains the following vertical performance verification:
• DC Gain Accuracy Test.
• Analog Bandwidth Test.
DC Gain Accuracy Test
Specifications
Equipment Required
CAUTION
Ensure that the input voltage to the oscilloscope never exceeds 300 Vrms.
Table 16 DC Gain Accuracy Specification
DC Gain Accuracy 2 mV/div to 5 mV/d: ±4.0% full scale
10 mV/div to 5 V/div: ±3.0% full scale
Full scale is defined as 8 vertical divisions. The major scale settings are 2 mV, 5 mV, 10 mV, 20 mV,
50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, and 5 V.
Table 17 Equipment Required for DC Gain Accuracy Test
Description Critical Specifications Recommended
Model/Part Numbers
Power Supply 0 V to 35 V DC; 10 mV resolution Agilent E3633A or
E3634A
Digital Multimeter DC voltage measurement accuracy better
than ±0.1% of reading
Agilent 34401A