User`s guide
6-25
Calibrating for Increased Measurement Accuracy
Enhanced Frequency Response Error Correction
18.To compute the error coefficients, press:
The analyzer displays the corrected measurement trace. The analyzer also shows the
notation Cor at the left of the screen, indicating that error correction is on.
NOTE You can save or store the measurement correction to use for later
measurements. Refer to Chapter 4 , “Printing, Plotting, and Saving
Measurement Results” for procedures.
19.This completes the enhanced response correction procedure. You can connect and
measure your device under test.
DONE ENH RESP CAL