User`s guide
1-12
Making Measurements
Using Display Functions
Viewing Both Primary Measurement Channels
In some cases, you may want to view more than one measured parameter at a time.
Simultaneous gain and phase measurements, for example, are useful in evaluating
stability in negative feedback ampliļ¬ers. You can easily make such measurements using
the dual channel display.
1. To see channels 1 and 2 in the same grid, press:
, set to ON, and
to 1X.
Figure 1-7 Example of Viewing Channel 1 and 2 Simultaneously
2. To view the measurements on separate graticules, press: Set to 2X. The
analyzer shows channel 1 on the upper half of the display and channel 2 on the lower
half of the display. The analyzer defaults to measuring S
11
on channel 1 and S
21
on
channel 2.
Display
DUAL | QUAD SETUP
DUAL CHAN on OFF
SPLIT DISP