User`s guide

2-35
Making Mixer Measurements (Option 089 Only)
Phase or Group Delay Measurements
1. Set the LO source to the desired CW frequency of 1000 MHz and power level to 13 dBm.
2. Initialize the analyzer by pressing .
3. Set the analyzer’s LO frequency to match the frequency of the LO source by pressing:
4. From the front panel of the analyzer, set the desired receiver frequency and source
output power by pressing:
( , 8722ES)
5. Connect the instruments as shown in Figure 2-25, placing a broadband "calibration"
mixer (ZFM-4) between PORT 1 and PORT 2.
CAUTION To prevent connector damage, use an adapter (part number 1250-1462) as a
connector saver for R CHANNEL IN.
Model Number Useful Frequency
Range
Group Delay
ANZAC
MCD-123
0.03 to 3 GHz 0.5 ns
Mini-Circuits
ZFM-4
dc to 1250 MHz 0.6 ns
Preset
System
INSTRUMENT MODE
FREQ OFFS MENU
LO FREQUENCY
1000
M/µ
Center 300 M/µ
Span 100 M/µ
Power 0
PWR RANGE MAN
x1 10 x1