User`s guide
6-20 User’s and Service Guide
Service
Assembly Replacement and Post-Repair Procedures
NOTE Special options are built to order, therefore long lead times may be
encountered when ordering replacement parts.
WARNING Some parts in the instrument have sharp edges. Work carefully
to avoid injury.
CAUTION Before replacing an assembly or board, inspect for obvious, easily
repaired defects such as bent pins on ICs or cold solder joints.
Connector Replacement
The 50 Ω 3.5 mm connectors are available separately as part number
5062-6618. It is possible to replace them in the field. An alternative to
replacing a damaged connector would be to replace just the center pin
components. The components for the assembly are:
When replacing just the pin and bead, a liquid thread-locking adhesive
such as part number 0470-1590 will be needed. Re-use any shims and
spacers from the connector being replaced. For best results, use a
connector gauge to verify pin depth. See Figure 4-2 on page 4-3 for
proper pin-depth. Add or subtract spacers as required. Spacers and
shims are also available from Agilent Technologies.
Component Part Number
3.5 mm Connector, RF (complete) 5062-6618
3.5 mm Pin and Bead Assembly (only) 5061-5394