User`s guide
User’s and Service Guide 3-9
Controlling the Test Set and Making Measurements
Calibrating the Test System
Calibrating the Test System
After the test set has warmed up for at least two hours, you should
calibrate before making any measurements. Refer to your network
analyzer user’s guide to determine the type of calibration appropriate
for the measurements you will be making.
You must calibrate each measurement path separately and store the
calibration as an instrument state in the network analyzer. Refer to
your network analyzer user’s guide for information on how to calibrate
and store instrument states.
Refer to Figure 3-2. In this example setup the following tests will be
made:
• Return loss on the DUT input and 2 output ports (A and B)
• Insertion loss (or gain) between the DUT input and port A
• Insertion loss (or gain) between the DUT input and port B
Figure 3-2 Calibrating the Test System
For the best accuracy, you should perform a full two-port calibration
between test set ports 1 and 2, and again between ports 1 and 3. As
mentioned before, you must save the calibrations as instrument states.
See your analyzer user’s guide for information on calibrations and
saving instrument states.