User`s guide
6 User’s and Service Guide
87050A Option K24 General Specifications
Environmental Requirements
Operating Environment
Indoor use only
Operating Temperature: 0 to 55 °C
(32 °F to 104 °F)
Storage Temperature: -40 °C to +70 °C (–40 °F to +158 °F)
Altitude: 10,000 feet (3,000 meters)
Enclosure Protection IP 2 0
Environmental Tests
The 87050A Option K24 complies with all applicable safety and regulatory requirements
for the intended location of use and have been evaluated to assure that they are consistent
with Agilent quality and reliability goals.
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient temperature
within the appropriate range. Air conditioning capacity must be consistent with the power
ratings given on Page 5.
Required Conditions for Accuracy Enhanced Measurement
Accuracy–enhanced (error–corrected) measurements require the ambient temperature of
the PNA and Test Set to be maintained within ± 1 °C of the ambient temperature at
calibration.
CAUTION This product is designed for use in Installation Category II, and Pollution
Degree 2.