Technical data
736 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
33 Obsolete and Discontinued Commands
:MEASure:VSTOp
(see page 790)
Command Syntax
:MEASure:VSTOp <vstop_argument>
<vstop_argument> ::= value for Y2 cursor
The :MEASure:VSTOp command moves the vertical marker 2 (Y2 cursor) to
the specified value corresponding to the selected source. The source can
be selected by the MARKer:X2Y2source command.
Query Syntax
:MEASure:VSTOp?
The :MEASure:VSTOp? query returns the current value of the Y2 cursor.
Return Format
<value><NL>
<value> ::= value of the Y2 cursor in NR3 format
See Also • "Introduction to :MARKer Commands" on page 318
• "Introduction to :MEASure Commands" on page 340
• ":MARKer:Y1Position" on page 327
• ":MARKer:Y2Position" on page 328
• ":MARKer:YDELta" on page 329
• ":MARKer:X2Y2source" on page 323
• ":MEASure:SOURce" on page 361
• ":MEASure:TDELta" on page 727
• ":MEASure:TSTArt" on page 729
NOTE
The short form of this command, VSTO, does not follow the defined Long Form to Short
Form Truncation Rules (see page 792). The normal short form, VST, would be the same for
both VSTArt and VSTOp, so sending VST for the VSTOp command produces an error.
NOTE
The :MEASure:VSTOp command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:Y2Position command (see page 328) instead.