Technical data
:WGEN Commands 31
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 667
:WGEN:FUNCtion
(see page 790)
Command Syntax
:WGEN:FUNCtion <signal>
<signal> ::= {SINusoid | SQUare | RAMP | PULSe | NOISe | DC}
The :WGEN:FUNCtion command selects the type of waveform:
Waveform
Type
Characteristics
SINusoid Use these commands to set the sine signal parameters:
• ":WGEN:FREQuency" on page 666
• ":WGEN:PERiod" on page 686
• ":WGEN:VOLTage" on page 688
• ":WGEN:VOLTage:OFFSet" on page 691
• ":WGEN:VOLTage:HIGH" on page 689
• ":WGEN:VOLTage:LOW" on page 690
The frequency can be adjusted from 100 mHz to 20 MHz.
SQUare Use these commands to set the square wave signal parameters:
• ":WGEN:FREQuency" on page 666
• ":WGEN:PERiod" on page 686
• ":WGEN:VOLTage" on page 688
• ":WGEN:VOLTage:OFFSet" on page 691
• ":WGEN:VOLTage:HIGH" on page 689
• ":WGEN:VOLTage:LOW" on page 690
• ":WGEN:FUNCtion:SQUare:DCYCle" on page 671
The frequency can be adjusted from 100 mHz to 10 MHz.
The duty cycle can be adjusted from 20% to 80%.
RAMP Use these commands to set the ramp signal parameters:
• ":WGEN:FREQuency" on page 666
• ":WGEN:PERiod" on page 686
• ":WGEN:VOLTage" on page 688
• ":WGEN:VOLTage:OFFSet" on page 691
• ":WGEN:VOLTage:HIGH" on page 689
• ":WGEN:VOLTage:LOW" on page 690
• ":WGEN:FUNCtion:RAMP:SYMMetry" on page 670
The frequency can be adjusted from 100 mHz to 100 kHz.
Symmetry represents the amount of time per cycle that the ramp waveform is
rising and can be adjusted from 0% to 100%.