Technical data
:TIMebase Commands 28
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 585
:TIMebase:WINDow:POSition
(see page 790)
Command Syntax
:TIMebase:WINDow:POSition <pos value>
<pos value> ::= time from the trigger event to the zoomed (delayed)
view reference point in NR3 format
The :TIMebase:WINDow:POSition command sets the horizontal position in
the zoomed (delayed) view of the main sweep. The main sweep range and
the main sweep horizontal position determine the range for this command.
The value for this command must keep the zoomed view window within
the main sweep range.
Query Syntax :TIMebase:WINDow:POSition?
The :TIMebase:WINDow:POSition? query returns the current horizontal
window position setting in the zoomed view.
Return Format
<value><NL>
<value> ::= position value in seconds
See Also • "Introduction to :TIMebase Commands" on page 578
• ":TIMebase:MODE" on page 579
• ":TIMebase:POSition" on page 580
• ":TIMebase:RANGe" on page 581
• ":TIMebase:SCALe" on page 583
• ":TIMebase:WINDow:RANGe" on page 586
• ":TIMebase:WINDow:SCALe" on page 587