Technical data

:SYSTem Commands 27
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 573
:SYSTem:PROTection:LOCK
(see page 790)
Command Syntax
:SYSTem:PROTection:LOCK <value>
<value> ::= {{1 | ON} | {0 | OFF}}
The :SYSTem:PROTection:LOCK command disables the fifty ohm impedance
setting for all analog channels.
Query Syntax
:SYSTem:PROTection:LOCK?
The :SYSTem:PROTection:LOCK? query returns the analog channel
protection lock status.
Return Format
<value><NL>
<value> ::= {1 | 0}
See Also "Introduction to :SYSTem Commands" on page 564