Technical data

:SEARch Commands 26
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 545
:SEARch:SERial:IIC:PATTern:ADDRess
(see page 790)
Command Syntax
:SEARch:SERial:IIC:PATTern:ADDRess <value>
<value> ::= integer or <string>
<string> ::= "0xnn" n ::= {0,..,9 | A,..,F}
The :SEARch:SERial:IIC:PATTern:ADDRess command specifies address
values when searching for IIC events.
To set don't care values, use the integer - 1.
Query Syntax
:SEARch:SERial:IIC:PATTern:ADDRess?
The :SEARch:SERial:IIC:PATTern:ADDRess? query returns the current
address value setting.
Return Format
<value><NL>
<value> ::= integer
See Also Chapter 26, “:SEARch Commands,” starting on page 531
":SEARch:SERial:IIC:MODE" on page 543