Technical data
540 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
26 :SEARch Commands
:SEARch:SERial:CAN:PATTern:ID
(see page 790)
Command Syntax
:SEARch:SERial:CAN:PATTern:ID <string>
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X}
for hexadecimal
The :SEARch:SERial:CAN:PATTern:ID command specifies the ID value when
searching for a CAN event.
The value can be a standard ID or an extended ID, depending on the
:SEARch:SERial:CAN:PATTern:ID:MODE command's setting.
Query Syntax
:SEARch:SERial:CAN:PATTern:ID?
The :SEARch:SERial:CAN:PATTern:ID? query returns the current ID value
setting.
Return Format
<string><NL>
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X}
for hexadecimal
See Also • Chapter 26, “:SEARch Commands,” starting on page 531
• ":SEARch:SERial:CAN:MODE" on page 537
• ":SEARch:SERial:CAN:PATTern:ID:MODE" on page 541