Technical data

:SBUS<n> Commands 25
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 529
:SBUS<n>:UART:TRIGger:TYPE
(see page 790)
Command Syntax
:SBUS<n>:UART:TRIGger:TYPE <value>
<value> ::= {RSTArt | RSTOp | RDATa | RD1 | RD0 | RDX | PARityerror
| TSTArt | TSTOp | TDATa | TD1 | TD0 | TDX}
The :SBUS<n>:UART:TRIGger:TYPE command selects the UART trigger type.
When one of the RD or TD types is selected, the
:SBUS<n>:UART:TRIGger:DATA and :SBUS<n>:UART:TRIGger:QUALifier
commands are used to specify the data value and comparison operator.
The RD1, RD0, RDX, TD1, TD0, and TDX types (for triggering on data and
alert bit values) are only valid when a 9- bit width has been selected.
Query Syntax
:SBUS<n>:UART:TRIGger:TYPE?
The :SBUS<n>:UART:TRIGger:TYPE? query returns the current UART
trigger data value.
Return Format
<value><NL>
<value> ::= {RSTA | RSTO | RDAT | RD1 | RD0 | RDX | PAR | TSTA |
TSTO | TDAT | TD1 | TD0 | TDX}
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:MODE" on page 598
":SBUS<n>:UART:TRIGger:DATA" on page 526
":SBUS<n>:UART:TRIGger:QUALifier" on page 528
":SBUS<n>:UART:WIDTh" on page 530