Technical data

:SBUS<n> Commands 25
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 487
:SBUS<n>:LIN:TRIGger
(see page 790)
Command Syntax
:SBUS<n>:LIN:TRIGger <condition>
<condition> ::= {SYNCbreak | ID | DATA}
The :SBUS<n>:LIN:TRIGger command sets the LIN trigger condition to be:
SYNCbreak — Sync Break.
ID — Frame ID.
Use the :SBUS<n>:LIN:TRIGger:ID command to specify the frame ID.
DATA Frame ID and Data.
Use the :SBUS<n>:LIN:TRIGger:ID command to specify the frame ID.
Use the :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth and
:SBUS<n>:LIN:TRIGger:PATTern:DATA commands to specify the data
string length and value.
Query Syntax :SBUS<n>:LIN:TRIGger?
The :SBUS<n>:LIN:TRIGger? query returns the current LIN trigger value.
Return Format
<condition><NL>
<condition> ::= {SYNC | ID | DATA}
Errors "- 241, Hardware missing" on page 753
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:MODE" on page 598
":SBUS<n>:LIN:TRIGger:ID" on page 488
":SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth" on page 491
":SBUS<n>:LIN:TRIGger:PATTern:DATA" on page 489
":SBUS<n>:LIN:SOURce" on page 484