450 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
:SBUS<n>:DISPlay
(see page 790)
Command Syntax
:SBUS<n>:DISPlay <display>
<display> ::= {{1 | ON} | {0 | OFF}}
The :SBUS<n>:DISPlay command turns displaying of the serial decode bus
on or off.
Query Syntax
:SBUS<n>:DISPlay?
The :SBUS<n>:DISPlay? query returns the current display setting of the
serial decode bus.
Return Format
<display><NL>
<display> ::= {0 | 1}
Errors • "- 241, Hardware missing" on page 753
See Also • "Introduction to :SBUS<n> Commands" on page 447
• ":CHANnel<n>:DISPlay" on page 222
• ":DIGital<d>:DISPlay" on page 245
• ":POD<n>:DISPlay" on page 412
• ":VIEW" on page 182
• ":BLANk" on page 158
• ":STATus" on page 179
NOTE
This command is only valid when a serial decode option has been licensed.
NOTE
Two I2S buses or two SPI buses cannot be decoded on both SBUS1 and SBUS2 at the same