Technical data

190 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
7 :ACQuire Commands
:ACQuire:SEGMented:COUNt
(see page 790)
Command Syntax
:ACQuire:SEGMented:COUNt <count>
<count> ::= an integer from 2 to 25 (w/100K memory) in NR1 format
The :ACQuire:SEGMented:COUNt command sets the number of memory
segments to acquire.
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command, and data is acquired using the :DIGitize,
:SINGle, or :RUN commands. The number of memory segments in the
current acquisition is returned by the :WAVeform:SEGMented:COUNt?
query.
The maximum number of segments may be limited by the memory depth
of your oscilloscope. For example, an oscilloscope with 100K memory
allows a maximum of 25 segments.
Query Syntax
:ACQuire:SEGMented:COUNt?
The :ACQuire:SEGMented:COUNt? query returns the current count setting.
Return Format
<count><NL>
<count> ::= an integer from 2 to 25 (w/100K memory) in NR1 format
See Also ":ACQuire:MODE" on page 187
":DIGitize" on page 159
":SINGle" on page 178
":RUN" on page 176
":WAVeform:SEGMented:COUNt" on page 647
":ACQuire:SEGMented:ANALyze" on page 189
"Introduction to :ACQuire Commands" on page 183
Example Code "Example Code" on page 191
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.