Technical data

:ACQuire Commands 7
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 189
:ACQuire:SEGMented:ANALyze
(see page 790)
Command Syntax
:ACQuire:SEGMented:ANALyze
This command calculates measurement statistics and/or infinite
persistence over all segments that have been acquired. It corresponds to
the front panel Analyze Segments softkey which appears in both the
Measurement Statistics and Segmented Memory Menus.
In order to use this command, the oscilloscope must be stopped and in
segmented acquisition mode, with either quick measurements or infinite
persistence on.
See Also ":ACQuire:MODE" on page 187
":ACQuire:SEGMented:COUNt" on page 190
"Introduction to :ACQuire Commands" on page 183
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.