Technical data

Root (:) Commands 6
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 175
:PRINt
(see page 790)
Command Syntax
:PRINt [<options>]
<options> ::= [<print option>][,..,<print option>]
<print option> ::= {COLor | GRAYscale | PRINter0 | PRINter1 | BMP8bit
| BMP | PNG | NOFactors | FACTors}
The <print option> parameter may be repeated up to 5 times.
The PRINt command formats the output according to the currently
selected format (device). If an option is not specified, the value selected in
the Print Config menu is used.
See Also "Introduction to Root (:) Commands" on page 150
"Introduction to :HARDcopy Commands" on page 296
":HARDcopy:FACTors" on page 299
":HARDcopy:GRAYscale" on page 722
":DISPlay:DATA" on page 258