Technical data
Root (:) Commands 6
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 167
:OPERegister:CONDition (Operation Status Condition Register)
(see page 790)
Query Syntax
:OPERegister:CONDition?
The :OPERegister:CONDition? query returns the integer value contained in
the Operation Status Condition Register.
Return Format
<value><NL>
<value> ::= integer in NR1 format.
See Also • "Introduction to Root (:) Commands" on page 150
• ":CHANnel<n>:PROTection" on page 232
• ":OPEE (Operation Status Enable Register)" on page 165
Table 53 Operation Status Condition Register
Bit Name Description When Set (1 = High = True), Indicates:
14-12 --- --- (Not used.)
11 OVLR Overload A 50Ω input overload has occurred.
10 --- --- (Not used.)
9 MTE Mask Test Event A mask test event has occurred.
8-6 --- --- (Not used.)
5Wait
Trig
Wait Trig The trigger is armed (set by the Trigger Armed
Event Register (TER)).
4 --- --- (Not used.)
3 Run Running The oscilloscope is running (not stopped).
2-0 --- --- (Not used.)
:DLW
7ULJ
5XQ
$UP