Technical data

Root (:) Commands 6
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 159
:DIGitize
(see page 790)
Command Syntax
:DIGitize [<source>[,..,<source>]]
<source> ::= {CHANnel<n> | FUNCtion | MATH | SBUS1}
for the DSO models
<source> ::= {CHANnel<n> | DIGital<d> | POD{1 | 2}
| BUS{1 | 2} | FUNCtion | MATH | SBUS1}
for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The <source> parameter may be repeated up to 5 times.
The :DIGitize command is a specialized RUN command. It causes the
instrument to acquire waveforms according to the settings of the :ACQuire
commands subsystem. When the acquisition is complete, the instrument is
stopped.
If no argument is given, :DIGitize acquires the channels currently
displayed. If no channels are displayed, all channels are acquired.
See Also "Introduction to Root (:) Commands" on page 150
":RUN" on page 176
":SINGle" on page 178
":STOP" on page 180
":TIMebase:MODE" on page 579
Chapter 7, “:ACQuire Commands,” starting on page 183
Chapter 30, “:WAVeform Commands,” starting on page 627
NOTE
The :DIGitize command is only executed when the :TIMebase:MODE is MAIN or WINDow.
NOTE
To halt a :DIGitize in progress, use the device clear command.
NOTE
MATH is an alias for FUNCtion.