User`s manual

Connectivity
Standard ports One USB 2.0 hi-speed device port on rear panel. Supports USBTMC protocol
Two USB 2.0 hi-speed host ports, front and rear panel
Supports memory devices, printers and keyboards
Optional ports GPIB, LAN, WVGA video out
General and environmental characteristics
Power line consumption 100 watts
Power voltage range 100-120V, 50/60/400 Hz; 100-240V, 50/60 Hz ±10% auto ranging
Temperature Operating: 0 to +55 °C
Non-operating: –30 to +71 °C
Humidity Operating: Up to 80% RH at or below +40 °C; up to 45% RH up to +50 °C
Non-operating: Up to 95% RH up to 40 °C; up to 45% RH up to 50 °C
Altitude Operating: up to 4,000 m, Non-operating 15,300 m
Electromagnetic compatibility Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS
Safety UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04
Dimensions 381 mm (15 in) W x 204 mm (8 in) H x 142 mm (5.6 in) D
Weight Net: 3.9 kg (8.5 lbs), shipping: 4.1 kg (9.0 lbs)
Nonvolatile storage
Reference waveform display 2 internal waveforms or USB thumb drive
Waveform storage Set up, .bmp, .png, .csv, ASCII, XY, reference waveforms, .alb, .bin, lister, mask , HDFS
Max USB ash drive size Supports industry standard ash drives
Set ups without USB ash drive 10 internal setups
Set ups with USB ash drive Limited by size of USB drive
InniiVision X-Series physical characteristics
Included standard with oscilloscope
Standard 5-year warranty* (90 days for unserialized accessories)
Standard secure erase
Standard Probe
N2862B 150 MHz 10:1 passive probe Standard one per channel for 70 and 100 MHz models
N2863B 300 MHz, 10:1 passive probe Standard one per channel for 200 MHz models
N6459-60001 8-channel logic probe and accessory kit Standard on MSO models or with DSOX2MSO upgrade
Built-in help language support for English, Japanese, simplied Chinese, traditional Chinese, Korean, German, French, Spanish, Russian,
Portuguese and Italian, Certicate of Calibration, Documentation CD
Interface language support GUI menus: English, Japanese, simplied Chinese, traditional Chinese, Korean, German, French,
Spanish, Russian, Portuguese, Thai, Polish and Italian
Localized power cord
*Applies to all orders on or after 1/1/2013.
For MET/CAL procedures, click on the Cal Labs solutions link below
Cal Labs Solutions
http://www.callabsolutions.com/products/Keysight/ These procedures are FREE to customers.
20 | Keysight | InniiVision 2000 X-Series Oscilloscopes - Data Sheet