User`s manual
Integrated Digital Voltmeter
Functions ACrms, DC, DCrms, Frequency
Resolution ACV/DCV: 3 digits Frequency: 5.5 digits
Measuring rate 100 times/ second
Autoranging Automatic adjustment of vertical amplication to maximize the dynamic range of measurements.
Range meter Graphical display of most recent measurement, plus extrema over the previous 3 seconds.
Measurement range (specications are typical)
Frequency Range Vertical Range Vertical Accuracy
ACRms 20 Hz - 100 KHz 100 MHz to 500 MHz: 1 mV/div to 5 V/
div** (1 MΩ and 50 Ohm)
1 GHz model: 1 mV/div to 5 V/div** (1
MΩ), 1mV/div to 1V/div (50 Ohm)
[
DC vertical gain accuracy + 0.5% full scale
DCRms 20 Hz - 100 KHz [DC vertical gain accuracy + DC vertical offset ac-
curacy + 0.25% full scale]
DC NA [DC vertical gain accuracy + DC vertical offset ac-
curacy + 0.25% full scale]
Frequency
counter
1Hz – BW of Scope <10 mV/div: greater of 1 div or 5 mV; ≥
10 mV/div: 0.6 div
25 ppm ± 5 ppm per year (aging)
WaveGen – built-in function generator (specications are typical) (Continued)
Modulation Modulation types: AM, FM, FSK
Carrier waveforms: sine, ramp
Modulation source: Internal (no external modulation capability)
AM:
Modulation waveform: sine, square, ramp
Modulation frequency (1Hz to 20 kHz)
Depth: 0% to 100%
FM:
Modulation: sine, square, ramp (1 Hz to 20 kHz)
Modulation frequency (1 Hz to 20 kHz)
Minimum carrier frequency: 10 kHz
Minimum deviation: 1 Hz
Maximum deviation: 100 kHz or (carrier frequency - 9 kHz), whichever is smaller.
FSK:
Modulation: 50% duty cycle square wave
FSK rate: 1 Hz to 20 kHz
Minimum carrier frequency: 10 kHz
Minimum hop frequency: 2 * FSK rate
Performance characteristics (continued)
19 | Keysight | InniiVision 2000 X-Series Oscilloscopes - Data Sheet